![R&S 1444.1190K02 User Manual Download Page 117](http://html1.mh-extra.com/html/rands/1444-1190k02/1444-1190k02_user-manual_774881117.webp)
Remote control commands
R&S
®
NRPxxP
117
User Manual 1179.5760.02 ─ 02
Usage:
Event
SYSTem:TRANsaction:END
Ends a series of settings.
Usage:
Event
SYSTem[:SENSor]:NAME
<sensorname>
Sets the name of the power sensor according to your requirements. The specified
name is displayed in the web user interface of the network power sensors.
The name that you specify here is independent from the hostname of the power sen-
sor. However, if you do not specify a name, the hostname is used as default.
Parameters:
<sensorname>
Example:
SYST:NAME "InputModule-X5"
SYSTem:VERSion?
Queries the SCPI version that the command set of the power sensor complies with.
Example:
SYST:VERS?
Query
1999.0
Response: SCPI version from 1999.
Usage:
Query only
8.13
Using the status register
Further information:
●
Chapter 10.2, "Status reporting system"
Contents:
General status register commands
....................................................................... 118
..................................................................................118
........................................................................................ 118
.................................................................................. 119
Controlling the negative transition part
..................................................................119
Controlling the positive transition part
................................................................... 119
Using the status register