Remote control commands
R&S
®
NRPxxP
71
User Manual 1179.5760.02 ─ 02
1.00
0.00
dBm
-30.00
0.00
1
2
Figure 8-6: PDF measurement parameter
1 = Average power
2 = Peak power
For information on querying the measurement results, see
8.6.6.1
Configuring the analysis window
Set the number of analysis window repetitions using
. For defining start, length and exclusion period, use the following
commands:
SENSe<Sensor>:STATistics[:EXCLude]:MID:OFFSet[:TIME]
................................................71
SENSe<Sensor>:STATistics[:EXCLude]:MID:TIME
..............................................................71
SENSe<Sensor>:STATistics:OFFSet:TIME
.........................................................................72
..................................................................................... 72
SENSe<Sensor>:STATistics[:EXCLude]:MID:OFFSet[:TIME]
<time>
Sets the start of an exclusion interval in the analysis window, referenced to the start of
the analysis window.
All other settings are the same as for the timeslot mode, see
Parameters:
<time>
Range:
0.0 to 0.3
*RST:
0.0
Default unit: s
SENSe<Sensor>:STATistics[:EXCLude]:MID:TIME
<time>
Sets the length of an exclusion interval in the analysis window.