Remote control commands
R&S
®
NRPxxP
58
User Manual 1179.5760.02 ─ 02
Example:
BUFF:SIZE 1
See
Chapter 9.2, "Performing a buffered continuous average
SENSe<Sensor>:[POWer:][AVG:]BUFFer:STATe
<state>
Enables or disables a buffered continuous average measurement. If enabled, the
power sensor collects all results generated by trigger events until the buffer is filled.
You can set the size of the buffer using
SENSe<Sensor>:[POWer:][AVG:]BUFFer:
.
Parameters:
<state>
ON | OFF
*RST:
OFF
Example:
BUFF:STAT OFF
8.6.2
Burst average measurement
The burst average mode is used to measure the average power of bursts. The integra-
tion time of a measurement is not predefined but determined by the power sensor with
the aid of a burst detector. The start of a burst is detected when the measurement sig-
nal rises above a set trigger level. The measurement ends when the signal drops
below a trigger threshold.
Power
Time
Trigger level
Dropout time
Dropout tolerance
Pulse interval
Figure 8-2: Burst average parameters
For information on querying the measurement results, see
Configuring the measurement modes