Remote control commands
R&S
®
NRPxxP
79
User Manual 1179.5760.02 ─ 02
Sets a time interval at the end of bursts that is excluded from the measurement. See
.
Parameters:
<exclude_stop>
Range:
0.0 to 51.6e-6
*RST:
0.0
Default unit: Seconds
8.7.4
Configuring corrections
It is possible to set some parameters that compensate for a change of the measured
signal due to fixed external influences.
.............................................................................................79
........................................................................................... 80
..............................................................................................81
Using the S-Parameters program
........................................................................... 83
8.7.4.1
Duty cycle corrections
The duty cycle is the percentage of one period during which the signal is active, when
pulse-modulated signals are corrected. The duty cycle is only evaluated in the continu-
ous average mode.
Remote commands:
SENSe<Sensor>:CORRection:DCYCle
..............................................................................79
SENSe<Sensor>:CORRection:DCYCle:STATe
....................................................................79
SENSe<Sensor>:CORRection:DCYCle
<duty_cycle>
Available in continuous average mode.
Sets the duty cycle for measuring pulse-modulated signals. The duty cycle defines the
percentage of one period during which the signal is active. If the duty cycle is enabled,
the R&S NRPxxP considers this percentage when calculating the signal pulse power
from the average power.
Parameters:
<duty_cycle>
Range:
0.001 to 100.00
*RST:
1.00
Default unit: Percent
SENSe<Sensor>:CORRection:DCYCle:STATe
<state>
Enables or disables the duty cycle correction for the measured value.
Parameters:
<state>
*RST:
OFF
Configuring basic measurement parameters