![R&S 1444.1190K02 User Manual Download Page 105](http://html1.mh-extra.com/html/rands/1444-1190k02/1444-1190k02_user-manual_774881105.webp)
Remote control commands
R&S
®
NRPxxP
105
User Manual 1179.5760.02 ─ 02
SENSe<Sensor>:TRACe:MEASurement:POWer:PULSe:DROop?
Queries the droop of the pulse. The droop is the difference between leading edge
amplitude, pulse top, and falling edge amplitude, as shown in
. The droop is
expressed in percentage of the pulse top amplitude.
The pulse droop measurement is used to test the stability of the output power of a
device, e.g. an amplifier or a transmitter.
Usage:
Query only
SENSe<Sensor>:TRACe:MEASurement:POWer:PULSe:TOP?
Queries the pulse top power level detected by the selected analysis algorithm,
SENSe<Sensor>:TRACe:MEASurement:ALGorithm
. This value is used as a refer-
ence (0 %) to determine other parameter values such as the rising or falling thresholds.
Usage:
Query only
SENSe<Sensor>:TRACe:MEASurement:POWer:REFerence?
Queries the power level at reference level. See
MEASurement:DEFine:DURation:REFerence
.
Usage:
Query only
SENSe<Sensor>:TRACe:MEASurement:TRANsition:NEGative:OVERshoot?
Queries the height of the local minimum before a rising edge, divided by the pulse
amplitude:
The search interval begins when the signal falls below the lower reference level set by
SENSe<Sensor>:TRACe:MEASurement:DEFine:TRANsition:LREFerence
and
ends half a period thereafter.
Usage:
Query only
SENSe<Sensor>:TRACe:MEASurement:TRANsition:POSitive:OVERshoot?
Queries the height of the local maximum before a falling edge, divided by the pulse
amplitude:
Querying measurement results