![R&S 1444.1190K02 User Manual Download Page 68](http://html1.mh-extra.com/html/rands/1444-1190k02/1444-1190k02_user-manual_774881068.webp)
Remote control commands
R&S
®
NRPxxP
68
User Manual 1179.5760.02 ─ 02
HISTogram
Analyzes the histogram of the trace data. The average of all
points representing the pulse top is taken as pulse top power.
Similarly, the pulse base is determined.
This algorithm is suitable for most pulse signals.
INTegration
Fits a rectangle pulse of the same energy into the pulse signal
as a reference and thus determines the pulse top power.
This algorithm is recommended for:
Pulse signals with modulation
If the pulse energy is considered; for example, if you compare
the measurement result to a measurement result of a thermal
power sensor.
PEAK
Assigns the pulse peak power to the pulse top power, and the
pulse minimum power to the pulse base power.
Suitable for distance-measuring equipment (DME) pulses.
*RST:
HISTogram
SENSe<Sensor>:TRACe:MEASurement:DEFine:DURation:REFerence
<reference>
Sets the pulse width, pulse start time and pulse stop time. Used for the measurement
of pulse duration and period.
Parameters:
<reference>
Specified in percent of the pulse amplitude.
Range:
0.0 to 100.0
*RST:
50.0
Default unit: PCT
SENSe<Sensor>:TRACe:MEASurement:DEFine:TRANsition:HREFerence
<hreference>
Sets the end of the rising edge and the start of the falling edge of the pulse. Used for
the measurement of the rise or fall time.
Parameters:
<hreference>
Specified in percent of the pulse amplitude.
Range:
0.0 to 100.0
*RST:
90.0
Default unit: PCT
SENSe<Sensor>:TRACe:MEASurement:DEFine:TRANsition:LREFerence
<lreference>
Sets the start of the rising edge and the start of the falling edge of the pulse. Used for
the measurement of the rise or fall time.
Configuring the measurement modes