![R&S 1444.1190K02 User Manual Download Page 95](http://html1.mh-extra.com/html/rands/1444-1190k02/1444-1190k02_user-manual_774881095.webp)
Remote control commands
R&S
®
NRPxxP
95
User Manual 1179.5760.02 ─ 02
32 | 64
32-bit or 64-bit
If you omit the length, the R&S NRPxxP sets the last used
length.
Example for
REAL,32
format:
#6768000....<binary float values>....
Example for
REAL,64
format:
#71536000....<binary float values>....
<data[,length]>
<ASCii[,0 to 12]>
ASCii
List of comma-separated, readable values.
[,0 to 12]
Defines the number of decimal places.
The reset value 0 does not restrict the number of decimal pla-
ces.
Example for
ASCii,4
format:
1.2938e-06, -4.7269e-11, ...
*RST:
ASCii,0
8.9
Querying measurement results
After the measurement, you can query the measurement results.
8.9.1
Continuous average measurement results
FETCh<Sensor>:ARRay[:POWer][:AVG]?
.......................................................................... 95
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
........................................................................95
SENSe<Sensor>:[POWer:][AVG:]BUFFer:DATA?
................................................................96
FETCh<Sensor>:ARRay[:POWer][:AVG]?
Queries the last valid measurement result of a buffered continuous average measure-
ment.
To configure the measurand, use
before the measurement is initi-
ated.
Usage:
Query only
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
Queries the last valid measurement result of the measurand that was configured
before.
To configure the measurand, use
before the measurement is initi-
ated.
Querying measurement results