22
Linearity
-
-
Linearity correction (read section 10.6).
Determine
-
-
Determining linearity correction points.
Delete
-
-
Deleting linearity correction.
Corrections
-
-
Entering corrections for linearity correction
points.
Information
-
-
Information (read-only).
Serial number
0
-
Preview of scale serial number.
Reading unit: range 1
0.001
-
Preview of unit value of range I.
Reading unit: range 2
0.001
-
Preview of unit value of range II.
Range
6.009
-
Preview of range I value.
Range 2
0.000
-
Preview of range II value.
Weighing firmware
version
-
-
Version of the weighing module software.
Divisions
-
-
Display of A/D converter divisions.
Set default
-
-
Restoring platform's factory settings.
*)
- For I-range scales set
0
,
**)
- For I- or II-range scales set
0
value.
The above table presents menu structure for one platform.
When more than one platforms are in use, enter <Factory
parameters> menu to see a list of all the platforms.
Each platform is defined by an analogous set of parameters
(see the above table).
10.4. Metrological Data
Enter
<
Global /
Top bar
– metrology>
submenu, set required
values:
PARAMETER
NAME
DEFAULT
VALUES
DESCRIPTION
Activation
No
Yes, No
Enabling/disabling display of metrological
data in the top bar of the home screen.
Scale type
-
PUE 7.1, WLY,
WPY, HRP
Selecting scale type, the scale type is
conditioned by an indicator type.
Scale name
-
-
Special name and/or name of RADWAG
distributor's device (6 characters maximum).
The name entered for this parameter
replaces the selected “Scale type” value.