  Set respective parameter values and return to weighing, saving changes. 

 
 

11.3.2. Defining Transmission Parameters for HRP Weighing Module 

 

 

In order to define transmission parameters of HRP weighing 
module,  press  an  internal  SW1  switch  on  the  main  board  
of the module. Follow HRP service manual. 

 

  Make  sure  that  communication  between  the  PUE  7.1  and  the  HRP  is 

established, enter factory parameters submenu of the PUE 7.1 indicator. 

  Go to a modified submenu of the weighing platform. 

  Enter 

„Miscellaneous / MW-MH: Communication”

  submenu,  list  of  the 

following parameters is displayed: 

 

NAME 

DEFAULT 

VALUES 

DESCRIPTION 

RS232 (1) 

57600 

1200

115200 

Baud  rate  of  RS232  (1)  port  of  HRP 
weighing module. 

RS485 

57600 

1200

115200 

Baud  rate  of  RS485  port  of  HRP 
weighing module. 

Address 

Setting address of HRP module. 

IP Address 

192.168.0.2 

Setting IP address of HRP module. 

Subnet mask 

255.255.255.0 

Setting subnet mask of HRP module. 

Default gateway 

192.168.0.1 

Setting default gateway of HRP module. 

Port 

4001 

Setting local port of HRP module. 

Timeout 

60 

0

300s 

Setting  time  interval  after  passage  
of 

which 

an 

inactive 

Ethernet 

connection gets broken. 

 

  Set respective parameter values and return to weighing, saving changes. 

 

 

In  order  to  save  changes  in  HRP  module  memory  it  is 
necessary to restart PUE 7.1 indicator. 

 

 

11.3.3. Defining Factory Settings for HRP Module 

 

 

In  order  to  determine  factory  settings  of  HRP  weighing 
module  press  an  internal  SW1  button on  the  main  board of 
the module. Follow HRP service manual. 

 

  Make  sure  that  communication  between  the  PUE  7.1  and  the  HRP  is 

established, enter factory parameters submenu of the PUE 7.1 indicator. 

Summary of Contents for PUE 7.1

Page 1: ......

Page 2: ...2 OCTOBER 2020...

Page 3: ...cator must not be operated in hazardous areas endangered with explosion of gases and in dusty environments Do not operate the touch panel using sharp edged tools knife screwdriver etc In case of damag...

Page 4: ...g Load Cell s Cable Shield 17 9 OPTIONAL EXTENSION MODULES 17 9 1 Additional Platform Module DP6 17 9 1 1 Technical Specifications 17 9 1 2 Colours of Weighing Platform s Signal Wires 18 10 FACTORY SE...

Page 5: ...of their elimination can only be realized with assistance of the manufacturer and user representatives C RADWAG does not bear any responsibility for damage or losses resulting from unauthorized or ina...

Page 6: ...f aliphatic hydrocarbons that do not dissolve plastics Before using the cleanser for all surfaces we recommend carrying out tests Do not use cleansers containing abrasive substances 3 2 Cleaning Stain...

Page 7: ...7 Overall dimensions PUE 7 1 panel version 4 2 Connectors 4 2 1 PUE 7 1 Connectors 1 Ethernet RJ45 connector 2 RS232 COM1 connector 3 USB connector 1 power supply socket 2 I O RS232 COM2 connector...

Page 8: ...connector 4 2 3 PUE 7 1P Cable Glands 1 cable gland for the power cord 2 1 2 platform cable gland 3 3 4 platform cable gland 4 2 4 RS232 and IN OUT Connector RS232 DB9 M connector male front Pin2 RxD...

Page 9: ...T1 Pin3 OUT2 Pin4 COMM Pin6 IN4 Pin7 IN3 Pin11 IN2 Pin12 IN1 Pin14 OUT4 Pin15 OUT3 4 3 1 Technical Specifications Output parameters Output quantity 4 Output type Solid state relay Cable cross section...

Page 10: ...ess to switch the scale ON OFF Press to zero the scale Press to tare the scale Press to send the weighing result to a printer or a computer Press to enter the menu function button Programmable key Pro...

Page 11: ...wer supply 5 V Load cell wiring 4 or 6 wires Shield Weighing platform quantity 2 maximum Communication interfaces 2xRS232 USB Ethernet 4IN 4OUT Wi Fi 2 4GHz b g n Standard None Multi range option YES...

Page 12: ...be found at the top of the operation panel Mind to keep the key pressed for about 0 7 s operating system and RADWAG program are loaded this is signalled with blinking of red ON LOAD diode Upon comple...

Page 13: ...the E2R SYSTEM is on Weighing device name Date and time 6 2 Weighing Result Window Weighing result window provides all weighing related data 6 3 Workspace The workspace is to be found underneath the w...

Page 14: ...s unmodified Press to add a database record Press to disable the selected record Press to log out Press to search a particular database record by date Press to search a particular database record by n...

Page 15: ...ns are automatically recorded upon return to the home screen To return to the home screen press key repeatedly keep pressing the key until you see the home screen press field located in the top bar th...

Page 16: ...d IN OUTPUT IN OUTPUT 5V INPUT AGND INPUT 8 2 4 Wire Load Cell Connection Connect 4 wire load cell to the main board following the diagram below 4 wire load cell connection Radwag A D converter board...

Page 17: ...A threaded pin connected with indicator s housing galvanically metal housing 9 OPTIONAL EXTENSION MODULES 9 1 Additional Platform Module DP6 The DP6 module expands indicator functionality with it an a...

Page 18: ...6 10 FACTORY SETUP To access and modify factory settings and parameters that are made available for a user run Factory Setup mode Running Factory Setup mode enables the technician user to define the s...

Page 19: ...nnecting with the indicator read section 11 Verified No No Yes Declaring whether the scale is verified or not Request modules pairing Yes No Yes Enabling disabling pairing of modules connecting with P...

Page 20: ...ty Enabling disabling accessibility of selected reports Database configuration Database setup submenu read PUE 7 1 Indicator Software Manual ct available No No Yes Enabling disabling metrological para...

Page 21: ...Switching point of range III Adjustment weight mass 2 Adjustment weight mass Adjustment unit kg g kg lb Adjustment unit g cor 1 Gravitational correction value Autozero range 0 25 0 1 5 Autozero range...

Page 22: ...form s factory settings For I range scales set 0 For I or II range scales set 0 value The above table presents menu structure for one platform When more than one platforms are in use enter Factory par...

Page 23: ...Upon completed procedure message Operation completed successfully is displayed Press button to confirm and return to weighing saving changes Adjustment of an additional platform is analogous 10 5 2 St...

Page 24: ...value specified for the first linearity correction point wait for a stable indication and press button Mass dialog box with an on screen keyboard opens declare the next linearity correction point Proc...

Page 25: ...value write sign before the number Press button to confirm list of linearity correction points is displayed again Proceed likewise for the next linearity correction point Return to weighing saving cha...

Page 26: ...basis of gravitational acceleration of the place where the calibration adjustment is carried out gravitational acceleration of the place of use Procedure Enter Global g cor submenu Select g cor metho...

Page 27: ...justed in RADOM press button to confirm the displayed value If the scale is adjusted elsewhere enter the value respective for a particular spot and press button to confirm Enter latitude degrees for p...

Page 28: ...1 indicator go to Global submenu Weighing module s transmission parameters baud rate computer port and PUE 7 1 indicator s transmission parameters must be compatible Configuration of transmission para...

Page 29: ...ice via Ethernet by default 4001 Reading unit range 1 0 001 0 001 50 Rounding of the last digit and position of a decimal point for weighing range I of the associated device Reading unit range 2 0 001...

Page 30: ...plement changes is displayed Press button to confirm and return to weighing saving changes next restart the device Enter factory parameters submenu of the PUE 7 1 indicator Go to a modified submenu of...

Page 31: ...set transmission parameters for MW 01 weighing module it is necessary to put a jumper onto the module s main board Procedure Make sure that communication between the PUE 7 1 and the MW 01 is establish...

Page 32: ...is displayed NAME VALUES DESCRIPTION Metrology Metrology related settings Reading unit range 1 0 001 50 Rounding of the last digit and position of a decimal point for weighing range I Reading unit ran...

Page 33: ...ns for determined linearity correction points Information Information read only Serial number Serial number of MW 01 module Weighing firmware version Preview of MW 01 module s software version Divisio...

Page 34: ...munication submenu list of the following parameters is displayed NAME DEFAULT VALUES DESCRIPTION Port None None Tcp RS232 1 RS232 2 RS485 Declaring port type of PUE 7 1 indicator for communication wit...

Page 35: ...ate of RS485 port of HRP weighing module Address 1 Setting address of HRP module IP Address 192 168 0 2 Setting IP address of HRP module Subnet mask 255 255 255 0 Setting subnet mask of HRP module Def...

Page 36: ...int of weighing range III Adjustment weight mass Adjustment weight mass Adjustment unit g kg lb Adjustment unit g cor Gravitational correction value Autozero range 0 1 5 Autozero range in divisions St...

Page 37: ...ion range Start mass value range in percent Automatic adjustment Submenu for setting the automatic internal adjustment Change position of internal adjustment weight Parameter allowing to reposition th...

Page 38: ...cient coefficient value is conditioned by temperature change and determined by the software after the end of compensation Adjustment temperature Value of divisions read from the temperature sensor dur...

Page 39: ...RP platform is automatically paired with the PUE 7 1 indicator message Modules paired is displayed Press button to confirm Serial number of the paired HRP weighing platform is displayed Pair modules f...

Page 40: ...40 Scale printer cable CITIZEN EPSON I O cable Scale Ethernet cable is a standard network cable terminated with RJ45 connectors on both ends...

Page 41: ...41...

Reviews: