11.3.1. Establishing Communication Between PUE 7.1 and HRP
Connection between PUE 7.1 indicator and HRP weighing module is carried
out via
RS232
,
RS485
or
Ethernet
interface.
Procedure:
Connect the HRP module to the PUE 7.1 indicator using a dedicated
RS232, RS485 or Ethernet cable (list of connection cables is to be found
in service manual of the HRP module).
Switch the PUE 7.1 indicator on and enter global parameters submenu.
Go to
<Weighing module type>
parameter and select the required
weighing module number.
Select
<HRP>
from the list. Message
<Restart to implement changes>
is displayed.
Press
button to confirm and return to weighing, saving changes,
next restart the device.
Enter factory parameters submenu of the PUE 7.1 indicator.
Go to a modified submenu of the weighing platform.
Enter
„Miscellaneous / PUE Y: Communication”
submenu, list of the
following parameters is displayed:
NAME
DEFAULT
VALUES
DESCRIPTION
Port *
None
None, Tcp,
RS232 (1),
RS232 (2), RS485
Declaring port type of PUE 7.1 indicator for
communication with HRP module.
Address
1
-
Declaring address of PUE 7.1 indicator for
communication with HRP module (default
address value in HRP module is
1
).
IP Address
0.0.0.0
-
Declaring IP address of PUE 7.1 indicator for
communication with HRP module via Ethernet.
IP port
0
-
Declaring IP port of PUE 7.1 indicator for
communication with HRP module via Ethernet.
*)
-
In case of connection via RS232, the default value of baud rate for HRP weighing module is
57600bit/s
.
Declaring RS232, RS485 or TCP port of PUE 7.1 indicator for
communication with HRP module requires respective
transmission parameters to be set for these ports.
HRP module's transmission parameters and transmission
parameters of PUE 7.1 indicator's ports must be compatible.
To set transmission parameters of PUE 7.1 indicator's ports
go to „SETUP/ Communication” submenu.