10.5. Factory Adjustment
10.5.1. External Adjustment
Enter
<
/
Factory/
Adjustment>
submenu and go to
<
Adjustment>
parameter.
Unload the platform.
Message
<Unload the platform>
is displayed.
Press
button. Start mass is determined, message
<Start mass
determination>
is displayed.
Upon completed start mass determination procedure, message
<Load
xxx>
is displayed (where: xxx
– the adjustment weight mass).
Load the platform with the required adjustment weight and press
button.
Upon
completed
procedure,
message
<Operation
completed
successfully>
is displayed.
Press
button to confirm and return to weighing, saving changes.
Adjustment of an additional platform is analogous.
10.5.2. Start Mass Determination
Enter
<
/
Factory /
Adjustment>
submenu and go to
<
Start mass determination>
parameter.
Unload the platform.
Message
<Unload the platform>
is displayed.
Press
button. Start mass is determined, message
<Start mass
determination>
is displayed.
Upon
completed
procedure,
message
<Operation
completed
successfully >
is displayed.
Press
button to confirm and return to weighing, saving changes.
Start mass determination for an additional platform is