31
Declaring RS232 or TCP port of PUE 7.1 indicator for
communication with MW-01 module requires respective
transmission parameters to be set for these ports.
MW-01
module's
transmission
parameters
and
transmission parameters of PUE 7.1 indicator's ports must
be compatible. To set transmission parameters of PUE 7.1
indicator's ports go to „SETUP/ Communication” submenu.
Set respective parameter values and return to weighing, saving changes.
11.2.2. Defining Transmission Parameters for MW-01 Weighing Module
In order to set transmission parameters for MW-01 weighing module it is
necessary to put a jumper onto the module's main board:
Procedure:
Make sure that communication between the PUE 7.1 and the MW-01 is
established, and enter factory parameters submenu of the PUE 7.1
indicator.
Go to a modified submenu of the weighing platform.
Enter
„Miscellaneous / MW-01: Communication”
submenu, list of the
following parameters is displayed:
NAME
DEFAULT
VALUES
DESCRIPTION
Address
1
-
Setting address of MW -01 module.
IP Address
192.168.0.2
-
Setting IP address of MW -01 module.
Subnet mask
255.255.255.0
-
Setting subnet mask of MW -01 module.
Default gateway
192.168.0.1
-
Setting default gateway of MW -01 module.
Port
4001
-
Setting local port of MW -01 module.