25
Select
<Corrections>
parameter. A list of determined linearity correction
points is displayed. The list contains previously declared mass values for
each correction point and deviations, given in reading units.
Select a respective entry,
<Mass>
dialog box with an on-screen keyboard,
providing a previously declared mass value is displayed.
Make sure that the weighing pan is unloaded,
and
pictograms
must be displayed, if not press
key to zero the scale.
Load the weighing pan with the declared mass and press
button.
<Correction>
dialog box with an on-screen keyboard opens automatically.
If the declared value differs from the displayed stabilised value then
a respective correction needs to be entered, use the on-screen keyboard.
If there is a need to enter negative value, write „-” sign before the number.
Press
button to confirm, list of linearity correction points is
displayed again.
Proceed likewise for the next linearity correction point.
Return to weighing, saving changes.
10.7. Deleting Linearity
Enter
<
Factory /
Linearity>
submenu.
Select
<Delete>
parameter. Message
<Delete?>
is displayed.
Press
button to confirm.
Return to weighing, saving changes.
10.8. Gravitational Correction
The function of gravitational correction compensates changes of earth gravity
being a result of different latitude and longitude. It enables correct calibration
of the weighing device away from the place of use. The value of gravitational
correction can be:
entered if known,
determined by gravitational acceleration method,
determined by latitude method.
10.8.1. Entering the Gravitational Correction Value
The gravitational correction value must be entered with reference to tables
prepared by „Radwag Balances and Scales” or calculated using the below
formula: