- 52 -
CAUTION
Function unavailable for verified balances.
•
P1.3
USER CALIB.
– User adjustment (with an external weight)
Adjustment with an external weight of any mass within balance range, however
not lower than 30% of Max range.
CAUTION
Function unavailable for verified balances.
•
P1.4
CALIB. TEST - Adjustment test
This fuction consists in comparing of an internal adjustment weight with its value
recorded in balance storage. The process is conducted automatically and its
result is shown on the display (if the balance is connected via one of the
interfaces with a computer or printer, the adjustement result will be printed out).
Press
ESC
to go back to the previous menu.
•
P1.5
AUTO. CALIB. T.
– Automatic adjustment (balances with an
internal adjustment)
Determining factor that is to be decisive when it comes to start of automatic
internal adjustment.
NONE
- neither time nor temperature is to trigger adjustement (adjustment
automatically turned off)
TEMP.
- adjustment triggered by change of temperature
TIME
- adjustment triggered by time interval set in
P1.6 AUTO CALIB. C.
BOTH
- adjustment triggered by time or temperature.