- 167 -
20.3.
Format of data sent on pressing PRINT button
CAUTION
Unstable measurement printout is not available for a verified
balance.
Printout format
1
2
3
4 - 12
13
14 - 16
17
18
st
ab
ilit
y
m
ar
ker
sp
ac
e
char
ac
ter
m
as
s
sp
ac
e
uni
t
CR
LF
stability marker - [space] – when stable
- [?] – when unstable
- [^] – if maximum range is exceeded
- [v] – if minimum range is exceeded
character
- [space] – for positive or [-] for negative values
mass
- 9 characters – aligned to the right
unit
- 3 characters – aligned to the left.
20.3.1.
Format of data sent as a response for commands generated
by a computer
After receiving a command, balance responds as follows first:
XX_A CR LF
command understood and in progress
XX_I
CR LF
command understood but not accessible at this moment
XX_^ CR LF
command understood but maximum range is exceeded
XX_v CR LF
command understood but minimum range is exceeded
XX_E CR LF
an error occurred on command carrying out (time limit
exceeded while waiting for stable measurement result),
time limit is balance’s characteristic parameter
XX
name of the command