- 133 -
•
Prepare a sample,
•
Initiate the process (like in case of solids density determination),
•
Insert settings for the process following the displayed messages,
•
Insert volume of the sinker used for measuring
•
On inserting the data, the program proceeds to weighing.
•
First, place the sinker on the hook (a sample mass determination in the
air), and confirm the indication when stable
•
Next, take the sinker out of the hook, put the beaker with liquid on the
beaker’s basis (the beaker cannot touch the hook), gently hang the
sinker on the hook providing that it is totally immersed in the tested
liquid (a mass sample determination in liquid) – and confirm the
indication when stable.