13.1. Internal adjustment
Adjustment process can be initiated automatically and manually.
Manual means of activating adjustment procedure is achieved by pressing
. System of automatic adjustment will carry out the process fully
automatically and will inform the user on successive process stages.
Cycle of automatic adjustment process:
−
Balance software detects the necessity of carrying out adjustment and
signals it by displaying a thermometer pictogram and <Cal> sign at the top
of the display.
−
−
2-minute long time interval starts, successive weighings may be performed
within this time,
−
As the 2-minute long time elapses, balance display indicates message
CAL_30
and starts counting down from 30..29..28 to 0 (indicated value is
the counter),
−
Balance user has 30 seconds to make a decision
In order to start adjustment, do not take any actions
In order to complete weighing procedure, press
Esc.
When
pressed, balance returns to weighing procedure and displays last
weighing result (the user may complete measurement series). In
about 5 minutes balance indicates
CAL 30
message again
.
−
The adjustment process can be postponed for multiple times, but it
needs to be pointed out that postponing of adjustment for a long time
may lead to larger errors of weighing process. The errors are the effect
of temperature changes and as a consequence changes of balance
sensitivity.