- 179 -
Mode name
– parameter, working mode name, inserted in between inverted
comas. The name takes form given on a particular balance display, it is
provided in a currently selected language.
n
– parameter, decimal value determining working mode number
n → 1 – Weighing
2 – Parts Counting
3 – Percent Weighing
4 – Dosing
5 – Formulas
6 – Animal Weighing
8 – Density of Solid Bodies
9 – Density of Liquids
10 – Peak Hold
11 – Totalizing
12 – Checkweighing
13 – Statistics
14 – Pipettes Calibration
CAUTION: Working modes numbering is identical for each kind of balance. The
numbers are assigned to working modes names. Some balances give only the
number as a response.
Example 1
Command:
OMI <CR><LF>
- give accessible working modes
Response:
OMI <CR><LF>
- accessible working modes are given;
2_”Parts Counting” <CR><LF>
name
4_”Dosing” <CR><LF>
12_”Checkweighing” <CR><LF>
OK <CR><LF>
- command carried out
Example 2
Command:
OMI <CR><LF>
- give accessible working modes
Response:
OMI <CR><LF>
- accessible working modes are given;
2<CR><LF>
number
4 <CR><LF>
12<CR><LF>
OK <CR><LF>