42
adjustment is automatically initiated if temperature changes more than 3
o
C
from last saved temperature.
-
adjustment triggered by elapsing time. It is possible to declare time
intervals which are a criteria for balance adjustment. Accessible settings
are: adjustment after 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12 hours since last
carried out adjustment procedure.
CAUTION
Declaring time interval for balance adjustment is available only in
non-verified balances. In case of verified balances the time interval
between following adjustments is set for 3 hours
12.2. Menu adjustments settings
P1 CAL
P1.1 iCAL
|
P1.2 uCAL
|
P1.3 tCAL
|
P1.4 ACAL
|
both
P1.5 CALt
|
1 h
P1.6 CALr
|
YES