34
10.5. Weighing mode RAPID
Weighing mode RAPD is located in menu parameter P3.rEAd and it offers two
options:
RAPID = NO
– balance operates with settings as set in menu AUE and CONF, i.e.
standard weighing.
RAPID = YES
– (default settings) balance operates with settings specific for
software mode “FILL”, and:
-
The value of CONF parameter is automatically set to: Fast+rel,
independently on previous setting of this parameter;
-
The value of AuE parameter is automatically set to: Standard,
independently on previous setting of this parameter;
-
The value of Stabilization time parameter is automatically decreased
(factory setting is decreased by 4). Such parameter setting speeds up the
time of stability pictogram
appearance on the display.
The values of AuE and CONF parameters are editable if mode rAPd is enabled for
setting shorter or longer measurement time.