19. ERROR MESSAGES
Er1 Hi
- mass out of range on balance start and adjustment
Er2 nuLL
- value from AD converter ≤ 0 (A/D ≤ 0)
Er3 FuL1
- values from A/D converter ≥ maximal converter range (A/D ≥ 0)
Er4 FuL2
– maximal capacity of the balance exceeded (over load)
Er5 rout
– mass value out of range (during adjustment, determining reference for
sample quantity and percent setup, etc.),
Er7 tout
- timeout error for zeroing and tarring
Er8 outr
- enter value (from keyboard level) out of range (setting thresholds / limits
values)
Er9 Lock
– function blocked (e.g. by LFT),
Er10 cal
– adjustment error (change of mass or incorrect mass of adjustment
weight / mass standard).