82
16.5.1. Format of data sent as response for commands generated
from a computer
On receipt of a command, balance responses with:
XX_A CR LF – command understood, in progress
XX_I CR LF
– command understood, but currently not available
XX _ ^ CR LF
– command understood, but max range is exceeded
XX _ v CR LF
– command understood, but min range is exceeded
XX _ E CR LF – error occurred while carrying out the command – time limit
exceeded while waiting for stable measurement result (time limit
is characteristic parameter of balance)
XX
– name of command
And followed by:
1 - 3
4
5
6
7
8 - 16
17
18 - 20
21
22
C
om
m
and
sp
ace
st
abi
lity
m
ar
ker
sp
ace
si
gn
m
ass
sp
ace
uni
t
CR
LF
Command
– 1 ÷ 3 characters
stability marker – [space] if stable
– [?] if unstable
– [^] if there is an error of range exceeding on +
– [v] if there is an error of range exceeding on -
sign
– [space] for positive values or [-] for negative values
mass
– 9 marks alignment to the right
unit
– 3 marks alignment to the left