5.  BASIC FUNCTIONS  

 

 

 

 

Functions related to measuring units 

-

 

Setting default measuring unit 

<StUn>

-

 

Weighing in grams [g],  

-

 

Weighing in milligrams [mg] (in balances with d=0,001 g) 

-

 

Weighing in kilograms [kg] (in balances with d=0,01 g) 

-

 

Weighing in [ct], 

-

 

Weighing  in  units  from  beyond 

SI 

(available  only  in  non-verified 

instruments: [lb], [oz], [ozt], [dwt], [t], [mom], [gn]. 

 

 

Functions related to working modes

 

-

 

Parts counting 

< PcS>

-

 

Checkweighing (C/-)

 <HiLo>

-

 

Percent setup  

<Perc A>, <Perc B>

-

 

Density determination of solids 

<d_Co> 

-

 

Density

 

determination of liquid 

<d_Li>

,

 

-

 

Totalizing 

<Add>

,

 

-

 

FiLL mode 

<FiLL>. 

 

 

Functions related to setting a balance to ambient conditions at a 
workstation 

-

 

Enabling / Disabling autozero system 

<Auto>

-

 

Setting parameter of measurement result averaging 

<AuE>

-

 

Value release 

<ConF>

-

 

Weighing mode 

<rAPd>

,  

-

 

Display backlight 

<bl>

-

 

Display brightness 

<bLbA>

-

 

Beep signal 

<bEEP>

 

 

 

Function relating to use of RS 232 interface 

-

 Setting baud rate 

<bAud>

-

 Continuous data transmission in basic measuring unit 

<CntA>

-

 Continuous data transmission in current measuring unit 

<Cntb>

-

 Automatic operation mode 

<rEPL>

 

Balance PS series can be used to weigh loads under the weighing platform – i.e. 

under hook weighing option.  

Balance software enables generating a report from adjustment process in a form of 

a printout containing modifiable data (see GLP procedures). 

 

Summary of Contents for PS 1000/C/2

Page 1: ...ENG Precision balances PS series M A N U F A C T U R E R O F E L E C T R O N I C W E I G H I N G I N S T R U M E N T S RADWAG 26 600 Radom Bracka 28 POLAND phone 48 48 38 48 800 fax 48 48 385 00 10 e...

Page 2: ...2 JUNE 2012...

Page 3: ...Cleaning 13 4 6 Plugging to mains 13 4 7 Connecting peripheral devices 13 5 BASIC FUNCTIONS 14 6 KEYBOARD 15 7 START UP 16 8 USER MENU 17 8 1 Moving through user menu 17 8 2 Moving through user menu 2...

Page 4: ...hing HiLo with reference to set standard 59 14 3 Percent setup 62 14 3 1 Reference mass determined by weighing 63 14 3 2 Reference mass inserted to balance memory by a user 64 14 4 Working mode FiLL 6...

Page 5: ...commands generated from a computer 82 17 UNDER HOOK WEIGHING 83 18 LIST OF COMMANDS COMPUTER BALANCE 84 19 ERROR MESSAGES 86 20 ADDITIONAL EQUIPMENT 87 20 1 Anti Vibration table 87 20 2 Rack for unde...

Page 6: ...6...

Page 7: ...2550ct 200 2000g 110g 210g 360g 510g 600g 750g 1000g Tare range 510g 2550ct 2000g 110g 210g 360g 510g 600g 750g 1000g Min load 0 02g 0 1ct 20mg 20mg 20mg 20mg 20mg 20mg 20mg 20mg Readability 0 001g 0...

Page 8: ...0500ct 1000g 1200g 2100g 3500g 4500g 6000g 8000g Tare range 2100g 10500ct 1000g 1200g 2100g 3500g 4500g 6000g 8000g Min load 0 5g 2 5ct 20mg 500mg 500mg 500mg 500mg 500mg 500mg Readability 0 01g 0 05c...

Page 9: ...ot place any magnetic materials on the weighing pan as this can cause damage of the measuring system of the instrument Be sure to avoid impact shock and overloading the balance in excess of the prescr...

Page 10: ...ff training This balance should only be operated and maintained by personnel who is trained and experienced in using this type of balances 3 TRANSPORT AND STORAGE 3 1 Delivery check Please check the p...

Page 11: ...s Take special safety measures when weighing magnetic objects as part of the balance is a strong magnet Should such loads be weighed use under hook weighing option which removes the weighed load from...

Page 12: ...e weighing pan is firmly placed on the rubber mandrels Assembling a balance with reading interval 1 mg Remove the adhesive tape protecting the mass spring on one of the 3 Assemble the weighing pan 2 o...

Page 13: ...can be connected to mains only with a power adapter that comes standard with the balance Nominal power supply of the power adapter specified on the data plate of the power adapter should be compatibl...

Page 14: ...e FiLL Functions related to setting a balance to ambient conditions at a workstation Enabling Disabling autozero system Auto Setting parameter of measurement result averaging AuE Value release ConF We...

Page 15: ...s working mode UNITS key changes measuring units PRINT ENTER key Sends current display status to a peripheral device PRINT or accepts selected value of a parameter or function ENTER ESC ZERO TARE key...

Page 16: ...For balances that were stored in much lower temperatures before plugging to mains e g during winter period thermal stabilization should last approximately 4 hours Correct operation of a balance is pos...

Page 17: ...menu Selecting group of parameters one by one downwards or changing parameter value by one value downwards Selecting group of parameters one by one upwards or changing parameter value by one value upw...

Page 18: ...me P2 6 PPrJ YES no printout of project name P2 7 PId YES no printout of balance s factory number P2 8 PFr YES no printout of frames P3 rEAd Main user parameters P3 1 AuE Stand Slouu FASt filtering le...

Page 19: ...b d_Co d_Li Add FiLL selection of working modes P6 2 PcS YES no parts counting P6 3 HiLo YES no checkweighing P6 4 PrcA YES no percent setup with reference to weighed standard P6 5 Prcb YES no percent...

Page 20: ...ving changes Press ESC TARE key for several times until the display shows command SAvE When display indicates the above command for saving changes press one of below options PRINT ENTER save changes o...

Page 21: ...ARE key Press Units key to set a measuring unit g mg ct and also if enabled in factory menu kg lb oz ozt dwt t mom gn place weighed object on balance s weighing pan and read result only on stabilizati...

Page 22: ...will be displayed on balance switching on In order to set the measuring unit go to submenu P5 Units press key for a few times The display indicates available measuring units in turns After selecting...

Page 23: ...th set basic measuring unit 9 2 Temporary measuring unit Function enables selecting a measuring unit which will be indicated next to mass readout during weighing process Set measuring unit will be in...

Page 24: ...3 Setting accessibility of balance working modes This group of parameters enables the user to declare functions working modes which are accessible on pressing of MODE key Enter group of parameters P6...

Page 25: ...modes are accessible during weighing function if all then select ALL and if only one from all available described below After setting the parameter press PRINT key Balance goes back to displaying subm...

Page 26: ...26 If for instance PcS mode is selected follow procedure in accordance with point 14 1 point to return to weighing Press Esc TARE key balance returns to weighing...

Page 27: ...selecting one of available working modes Balance user can enable or disable modes which are not used in the weighing process This provides quicker access to modes used on a balance On mode available...

Page 28: ...eeded On switching to weighing with the accuracy of the II weighing range the display signals it by indicating II pictogram on its right side and an additional marker of last but one digit of the weig...

Page 29: ...utozero settings last digit visibility These parameters are set in group P3 rEAd 10 1 Setting level of filtering Procedure using UNITS key to select desired value of filter settings StAnd normal norma...

Page 30: ...30 The higher filtering setting the longer weighing time...

Page 31: ...to set balance to present conditions it is necessary to determine the value release parameter as FAST_rEL Fast or rEL Depending on selected option weighing time will be shorter or longer Fast_rEL fast...

Page 32: ...y less than declared AUTOZERO range e g one division balance will zero automatically display marker of stable measurement result and precise zero marker If AUTOZERO function is enabled then each weigh...

Page 33: ...ng of the last digit on the display Function enables blanking of the last displayed digit ALuuAYS always nEuEr never uu_StAb when measurement result is stabilized Return to weighing see point 8 2 retu...

Page 34: ...tically set to Fast rel independently on previous setting of this parameter The value of AuE parameter is automatically set to Standard independently on previous setting of this parameter The value of...

Page 35: ...rameters which enable influencing its functionality The parameters are contained in group P7 othEr and they are backlight and beep signal and printout of all balance parameters through RS 232 port In...

Page 36: ...Return to weighing see 8 2 return to weighing CAUTION Balance software has implemented option of automatic backlight switch off bl Aut In case of this setting the backlight is switched automatically...

Page 37: ...37 100 maximum brightness of display backlight 20 minimum brightness of display backlight no backlight off...

Page 38: ...38 11 3 Beep sound reaction on pressing function key OFF signal of pressing function key disabled On signal of pressing function key enabled Return to weighing see 8 2 return to weighing...

Page 39: ...f all balance parameters through RS 232 port After pressing right arrow navigating key balance parameters are sent to a peripheral device connected to balance s RS 232 port Return to weighing see 8 2...

Page 40: ...anges more than 3 C Types of adjustment Internal automatic adjustment triggered by temperature change triggered by elapsing time Manual internal adjustment initiated from balance s keyboard Adjustment...

Page 41: ...value is the counter balance user has 30 seconds to make a decision in order to start adjustment do not take any actions in order to complete weighing procedure press ESC TARE key If pressed balance r...

Page 42: ...ent Accessible settings are adjustment after 1 2 3 4 5 6 7 8 9 10 11 12 hours since last carried out adjustment procedure CAUTION Declaring time interval for balance adjustment is available only in no...

Page 43: ...be carried out automatically but measuring results may differ by the amount that was present on balance weighing pan or CAL function key P1 2 ECAL External adjustment with an external weight adjustmen...

Page 44: ...rmination of factor which condirions start of automatic internal adjustment nonE none of factors will cause start of adjustment tEmP adjustment triggered by change of temperature timE adjustment trigg...

Page 45: ...automatic adjustment process is initiated P1 6 CALr printout of adjustment report Printout activated on completion of automatic adjustment procedure balances with internal adjustment balances with ex...

Page 46: ...ght with its value stored in balance s memory This process is automatic and its result is shown on the display if a balance is connected to a computer or a printer through RS 232 socket data on adjust...

Page 47: ...balance s memory during adjustment process This process is automatic and its result is shown on the display if a balance is connected to a computer or a printer through RS 232 socket data on adjustme...

Page 48: ...th any load On completion of adjustment process balance saves its result in memory and returns to weighing mode CAUTION Press ESC Tare key to abort adjustment process If during adjustment process weig...

Page 49: ...d ordering unloading of weighing pan unloAd weighing pan should be empty When weighing pan is unloaded press PRINT key The balance determines mass of empty pan Place an external adjustment weight with...

Page 50: ...set for YES then a report is generated and sent automatically Content of report from adjustment process depends on settings of GLP parameters Any option in the GLP submenu which attribute is YES is in...

Page 51: ...rintout activated after a measurement P2 GLP P2 1 uSr P2 2 PrJ P2 3 Ptin YES P2 4 PdAt YES P2 5 PuS YES P2 6 PPrJ YES P2 7 PId YES P2 8 PFrn YES P2 1 USr The parameter enables determining name of a op...

Page 52: ...52 Accessible characters and their equivalents displayed by the balance An instance of a user name inserted to a balance using capital letters takes below form...

Page 53: ...nce of project name inserted to a balance using capital letters takes below form PRRRADWAG An instance of project name inserted to a balance using small letters takes below form prrradwag P2 3 Ptin Op...

Page 54: ...tout P2 7 Pld Option determining presence of balance factory number on a printout P2 8 PFr Option enabling printing frames on a printout For above parameters please select one of two available setting...

Page 55: ...mass determined by a user PrcB Density determination of solids d_Co Density determination of liquid d_Li Totalizing Add Working mode Fill FiLL Press MODE key Pressing MODE key causes display of the f...

Page 56: ...standard version is equipped with an option for counting parts and small objects of the same unit mass Parts counting mode does not cooperate with other working modes of balance PS series Enable PcS...

Page 57: ...ENTER PRINT key and as the display indicates value 20 proceed as shown on below figure If option FrEE sample quantity is declared by a user is selected insert sample quantity used for determining mass...

Page 58: ...ndicates a command LoAd load the weighing pan with as many details as inserted while determining sample quantity Press PRINT key balance indicates sample s quantity pcs pictogram is visible on the dis...

Page 59: ...ntil stable measurement pictogram is displayed Only if visible on balance s display it is allowed to accept the declared quantity of parts by pressing PRINT key Otherwise balance will not accept the m...

Page 60: ...60 During setting limit values the following relations occur...

Page 61: ...61 SET LOW LIMIT LOW THRESHOLD digit selection set digit value accept inserted digit value SET HIGH LIMIT HIGH THRESHOLD digit selection set digit value accept inserted digit value...

Page 62: ...ill indicate error message and return to weighing Return to weighing 14 3 Percent setup Balance software enables controlling deviation in of weighed loads with determined reference mass Mass of a refe...

Page 63: ...place a load which mass will be accepted as the reference press PRINT to accept this working mode after few seconds indication 100 00 will be displayed From now on the display will not indicate mass...

Page 64: ...g function keys Select a digit Select digit value Determine the value of reference mass insert it to balance memory and accept it by pressing PRINT key display will indicate 0 00 From now on the displ...

Page 65: ...settings as default and ensuring quicker operation The settings automatically decrease Autozero range to 1 reading interval Mode activation is signaled by a black triangle pictogram visible on displa...

Page 66: ...e balance operates in accordance with the settings of AUE and CONF options The options enable setting shorter or longer measurement time Disabling FiLL mode On disabling FiLL mode the measurement are...

Page 67: ...n assemble the weighing pan with stand 1 and place beaker s basis 2 on it Components of the density kit 1 Weighing pan with stand 7 Thermometer 2 Bottom weighing pan of density determining kit for sol...

Page 68: ...of the density kit and then weighing the same sample in liquid on the bottom weighing pan 2 of the density kit As the sample is weighed in liquid the result of density determination is automatically...

Page 69: ...totalizes mass of the mixture and exposes its total mass on the display During totalizing process mass of each added component is sent through RS232 interface to a connected computer of a printer e g...

Page 70: ...70...

Page 71: ...setting speed of transmission P4 2 CntA continuous data transmission in basic measuring unit P4 3 CntB continuous data transmission in current measuring unit P4 4 rEPl setting operation mode for RS 2...

Page 72: ...72 15 1 Baud rate Select required baud rate value 2400 bit s 4800 bit s 9600 bit s 19200 bit s Return to weighing see point 8 2 return to weighing...

Page 73: ...2 Continuous transmission 15 2 1 Continuous transmission in basic measuring unit no continuous transmission disabled YES continuous transmission enabled Return to weighing see point 8 2 return to wei...

Page 74: ...74 15 2 2 Continuous transmission in current measuring unit no continuous transmission disabled YES continuous transmission enabled Return to weighing see point 8 2 return to weighing...

Page 75: ...es place according to the following scheme press Esc TARE key to zero the balance display will indicate pictograms of stable measurement and precise zero place the load balance will send first stable...

Page 76: ...15 4 Determination of data type sent through RS 232 output no sending stable or temporary unstable weighing result YES sending stable weighing result Return to weighing see point 8 2 return to weighi...

Page 77: ...tting automatic operation While in automatic operation data will not be transmitted through RS 232 to a connected computer or printer until mass reading is above Lo value Press key to select a digit w...

Page 78: ...value indicated on balance display can be sent through RS 232 output to a peripheral device in one of four accessible ways manually on pressing PRINT key automatically on stabilization of weighing res...

Page 79: ...79 16 1 Cross section through connecting conductors BALANCE COMPUTER 2 RxD 3 TxD 3 TxD 2 RxD 4 DTR 6 DSR 5 GND 5 GND 6 DSR 6 DTR 7 RTS 8 CTS 8 CTS 7 RTS...

Page 80: ...c printer Kafka SQS On connecting a statistic printer KAFKA SQ S to a balance it is possible to carry out statistics from completed measurements An instance of a printout including statistics from ser...

Page 81: ...urement or instantaneous mass indication will be sent CAUTION For verified balances printout of instantaneous measurement data is blocked Printout format 1 2 3 4 12 13 14 16 17 18 stability marker spa...

Page 82: ...curred while carrying out the command time limit exceeded while waiting for stable measurement result time limit is characteristic parameter of balance XX name of command And followed by 1 3 4 5 6 7 8...

Page 83: ...ension place for hook visible in the hole the suspension is installed permanently to balance mechanism In the hole install the hook for under hook weighing the hook is standard equipment of a balance...

Page 84: ...UNIT Command SUI CR LF send result in current unit now do not wait for stable reading Function CONTINUOUS TRANSMISSION IN BASIC MEASURING UNIT Command C1 CR LF start continuous transmission in basic u...

Page 85: ...85 should be omitted as they are added only for the purpose of proper legibility...

Page 86: ...the balance exceeded over load Er5 rout mass value out of range during adjustment determining reference for sample quantity and percent setup etc Er7 tout timeout error for zeroing and tarring Er8 ou...

Page 87: ...ing density determination of substances Racks are also used for measuring absorptiveness of substances e g foamed polystyrene The rack is made of mild steel Rack height is 330mm 20 3 Density determina...

Page 88: ...ords RAD KEY 2000 implemets a set of commands for controlling a balance e g tarring start of continuous transmission cyclic commands set to a balance REC FS 2000 application for control of formulas It...

Page 89: ...89 MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG WAGI ELEKTRONICZNE 26 600 Radom 28 Bracka POLAND phone 48 48 38 48 800 fax 48 48 385 00 10 e mail export radwag com www radwag com...

Page 90: ...90...

Reviews: