14.3.1. Reference mass determined by weighing
•
Enter
PrcA
mode
•
on balance weighing pan place a load which mass will be accepted as
the reference
•
press
to accept this working mode
•
after few seconds indication
100,00%
will be displayed.
From now on the display will not indicate mass of weighed load but deviation of
mass placed on the pan in relation to the reference (in %).