28
Return to weighing
(see point 8.2. – return to weighing).
9.4. Dual range balance (PS 200/2000/C)
Balance PS 200/2000/C series is a dual range weighing instrument. The accuracy
of the I weighing range is d1 = 0,001 g, and of the II weighing range it is d2 = 0,01
g.
Transfer from weighing with the accuracy of the I weighing range to weighing with
the accuracy of the II weighing range takes place automatically on exceeding Max1
point at 200 g (with no operator activity needed). On switching to weighing with the
accuracy of the II weighing range, the display signals it by indicating II pictogram on
its right side, and an additional marker of last but one digit of the weighing result.
From now on the balance weighs mass with the accuracy of the II weighing range.
In order to return to weighing in the accuracy of the I weighing range:
•
Take the weighed load off the weighing pan
•
As the indication returns to zero and pictograms
→
0
←
and
are lit
press
→
0/T
←
key on balance’s keyboard.
The balance returns to weighing with the accuracy of the I weighing range, and II
weighing range pictogram and marker of the last but one digit are blanked.