- 38 -
8.7. USER-DEFINED UNIT
You may declare two units. Displayed value of a user-defined unit is a multiplication of measured
mass value and a coefficient specified for the particular user-defined unit. The units can be freely
named with use of 3 characters maximum. By default the names are displayed as
[u1]
– user unit 1,
and
[u2]
– user unit 2.
8.8. BALANCE ZEROING
Zeroing is a function allowing to zero mass indication. In order to zero mass indication, press
button. Mass indication of zero value shall be displayed together with precise zero
and stability
markers.
Zeroing process is an equivalent for determining new zero point, recognized by the balance as
precise zero. Zeroing is possible only for stable status of display indication.
CAUTION!
Zeroing the display indication is possible only within ±2% range of instrument’s maximum capacity. If the
zeroed value is above ±2% of the maximumcapacity, then the software indicates a respective error message.
8.9. BALANCE TARING
Taring is a function allowing to determine net weight of a measured object. In order to determine net
weight of the object, place object’s container (packaging) on the weighing pan, and on stabilization of
measurement result press
key. The display indicates mass equal zero and symbols: Net and
. On taking off the weighed load and its packaging from the weighing pan, the display indicates
sum of total tared mass with minus sign.
Summary of Contents for AS X2 Series
Page 7: ...7 1 GENERAL INFORMATION 1 1 DIMENSIONS AS X2 series PS X2 series...
Page 8: ...8 APP X2 series WLC X2 series...
Page 96: ...96 Status Completed Signature...