8.15. COOPERATION WITH TITRATORS
In order to provide correct cooperation with Titrators, go to settings of standard printout content and
set <Mass value for a titrator> parameter to <Yes> value. With this, other variables for the printout
are turned off.
With <Yes> value on, a respective pictogram is displayed in the top bar of a home screen. The
pictogram informs on special format of mass printout, tolerable by TITRATORS.
8.16. DUAL RANGE BALANCES AND WEIGHING
(refers PS 200/2000.X2 and WLC 1/10.X2 balances)
An example for PS 200/2000.X2 balance
PS 200/2000.X2
balance is a dual range weighing device with readability for
I weighing range
d
1
=
0.001g, and d
2
= 0.01g for
II weighing range
.
Transition from weighing with accuracy of the
I weighing range
to weighing with accuracy of the
II
weighing range
takes place automatically on exceeding Max1 200g. On entering the accuracy of the
II weighing range, the display indicates
II
pictogram on its right side, last but one digit turns to grey
colour.
From now on the balance continues weighing in the II weighing range.