- 41 -
Filter level setting
Filter settings adjustment depends on the working environment. For the best possible conditions the
filter can work in a very fast mode (V.FAST value for Filter parameter); however, if the conditions are
poor (shakes, drafts), the filter should be set to slow or very slow option (SLOW or V. SLOW value for
Filter parameter). The effectiveness of the filter is different throughout the weighing range. The filter
works slower when “approaching” the weighed mass, it works more intensively for weighed mass
within the set range of the filter (parameter for setting filter range is accessible only from the service
menu – the user does not have any access to it).
Depending on the filter, the weighing time is shorter (V.FAST and FAST) or longer (SLOW and V.
SLOW).
CAUTION!
The higher filter level, the longer weighing time.
Value release
Since ambient conditions at a workplace vary, it is necessary to determine the value release
parameter in the most preferable way enabling balance adaptation, parameter options are:
FAST.+REL., FAST or RELIABLE. Depending on the selected option, weighing time is either shorter
or longer.
Autozero function
The software features an autozero function (Auto) ensuring precise mass indication. This function
automatically controls and corrects zero indication. When Autozero is enabled, it compares balance
indications at declared time interval e.g. 1s, provided that weighing pan is unloaded and display
indication is close to zero. If results vary less than declared AUTOZERO range e.g. one division,
Summary of Contents for AS X2 Series
Page 7: ...7 1 GENERAL INFORMATION 1 1 DIMENSIONS AS X2 series PS X2 series...
Page 8: ...8 APP X2 series WLC X2 series...
Page 96: ...96 Status Completed Signature...