- 13 -
•
remove tape protecting the grounding spring, located on one of the rubber mandrels (1)
•
install components following the above diagram:
•
weighing pan (2),
•
glass draft shield (3).
WLC X2, 128x128 mm
WLC X2, 195x195 mm
•
remove tape protecting the grounding spring, located on one of the rubber mandrels (1)
•
install components following the above diagram:
•
weighing pan (2),
•
glass draft shield (3).
APP balance
APP 10 balances are equipped with 2 weighing pans which can be used in turns. It must be
remembered that balance activation is only possible with one weighing pan assembled at a time.
Attempt of activating the balance with two weighing pans assembled simultaneously will result in
display of „– LH –„ error. In such case one of the pans needs to be removed.
Remove the weighing pan, next remove a transport lock (if
installed).
Undo bolt (1) installed to block adjustment mechanism.
Install a weighing pan (for APP 10 install one of the pans).
Make sure the weighing pan is placed properly on shock
absorbers: the weighing pan cannot touch the housing and
should be stable.