- 115 -
•
set balance parameters for cooperation with barcode reader:
< PORT> - selection of the port, to which the barcode reader is to be connected
accessible options: NONE, COM 1, COM 2
25.6. EXTERNAL BUTTONS
Submenu <EXTERNAL BUTTONS> comprises settings that enable the balance to cooperate with
external buttons: TARE and PRINT.
Procedure:
•
enter <P4 DEVICES> menu
•
enter menu group <EXTERNAL BUTTONS>
•
run the buttons
<TARE> set to <YES> value
< PRINT> set to <YES> value
•
exit the balance menu
CAUTION:
The balance program enables cooperation with one or both buttons. If you need to connect both external
buttons, first connect a several electrical outlet to COM 2 port, next connect buttons TARE and PRINT to the
electrical outlet. Connect the printer or terminal to COM 1 or to the electrical outlet (determine the PRINTER-
BALANCE transfer perameters). Every time the user presses TARE and PRINT external button, the balance
reacts as if TARE and PRINT buttons of the balance keyboard were pressed.
If you need to use only one button, connect it directly to COM 2 port or use the electrical outlet.
In order to ensure the proper cooperation, you need to remember:
•
to connect additional buttons to COM 2 port,
•
to connect the buttons to appropriate sockets of the electrical outlet if in use (see descriptions above),
•
to activate the buttons in the balance parameters (see description above),
•
to disconnect other external devices (additional display or barcode reader) on COM2 port (for these devices it should
be <NONE>),
•
to set the port for <PRINTER> to <COM 2> value if the printer is connected to the electrical outlet (CPU socket).
Set of external buttons TARE and PRINT.
Standard balance equipment does not include the set of external buttons.
25.7. AMBIENT CONDITIONS MODULE
Ambient Conditions module can be connected to the balance via COM1 or COM2 port. In order to
provide correct cooperation enter connected module address and baud rate for the port (port
settings) to which the ambient conditions module is connected (the address and baud rate data is to
be found on an ambient conditions data plate).
Summary of Contents for AS X2 Series