Programming Reference 4-26
•
<freq> sets the output frequency in units of Hertz
•
<ampl> sets the output amplitude in units of Volts
•
<offs> sets the output offset in units of Volts
•
<duty_cycle> sets the squarewave duty cycle in units of %
Parameter type
Numeric
Parameter range
<freq>
100e-6 to 50e6
<ampl>
80e-3 to 8
<offs>
-3.6 to +3.6
<duty_cycle> 1 to 99
APPLy:SQUare?
Response
The 3155 returns <freq>,<ampl>,<offs>,<duty_cycle>. Default values are
1e6,5,0,50
APPLy:PULSe{<freq>,<ampl>,<offs>,<delay>,<rise>,<high>,<fall>}
Purpose
This command is a high-level command that programs the 3155 to output pulse waveform along with its as-
sociated parameters
•
<freq> sets the output frequency in units of Hertz
•
<ampl> sets the output amplitude in units of Volts
•
<offs> sets the output offset in units of Volts
•
<delay> sets the delay to the pulse rise transition in units of %
•
<rise> sets the pulse rise time in units of %
•
<high> sets the pulse high time in units of %
•
<fall> sets the pulse fall time in units of %
Parameter type
Numeric
Parameter range
<freq>
100e-6 to 50e6
<ampl>
80e-3 to 8
<offs>
-3.6 to +3.6
<delay>
0 to 99.9
<rise>
0 to 99.9
<high>
0 to 99.9
<fall>
0 to 99.9
APPLy:PULSe?
Response
The 3155 returns <freq>, <ampl>,<offs>,<delay>,<rise>,<high>,<fall>.
Default values are
1e6,5,0,10,10,10,10
Summary of Contents for 3155
Page 4: ...This page was left intentionally blank...
Page 14: ...User Manual 3155 1 2 Overview Figure 1 1 3155 Series...
Page 41: ...User Manual 3155 Installation 2 15 Figure 2 15 Device Manager...
Page 44: ...User Manual 3155 2 18 Installation Figure 2 19 Customer Information Step...
Page 110: ...User Manual 3155 3 58 ArbConnection This page was left intentionally blank...
Page 150: ...User Manual 3155 Programming Reference 4 40...
Page 184: ...User Manual 3155 Programming Reference 4 74 This page was left intentionally blank...
Page 202: ...User Manual 3155 Specifications A 8 This page was left intentionally blank...