3155 User Manual
Performance Verification 5-3
Frequency Accu-
racy
Accuracy Specifications
:
0.01% of reading to 50.00 MHz
Equipment
:
Counter
1.
Configure the Model 3155 as follows:
Function: Square
Frequency:
10 Hz
Amplitude:
1 V
Remote Commands: :RES
:OUTP ON
:FUNC:SHAP SQU
:FREQ 10
:VOLT 1
2.
Set the counter to frequency measurement.
3.
Connect the Model 3155 output to the counter’s input.
Change the frequency setting as required for the test
and verify the reading on the counter display as follows:
3155 SETTING
COUNTER READING
10.00000 Hz
9.999000 Hz - 10.00100 Hz
100.0000 Hz
99.99000 Hz - 100.0100 Hz
1.000000 kHz 999.9000 Hz - 1.000100 kHz
10.00000 kHz
9.999000 kHz - 10.00100 kHz
100.0000 kHz
99.99000 kHz - 100.0100 kHz
1.000000 kHz
999.9000 kHz - 1.000100 kHz
10.00000 MHz
9.999000 MHz - 10.00100 MHz
50.00000 MHz
49.99500 MHz - 50.00500 MHz
Amplitude Accu-
racy
Accuracy Specifications (1 kHz):
±
(1% of r1 mV), to 799.9 mV
±
(1% of r10 mV), to 8 V
Equipment
:
DMM, 50O feedthrough termination
1.
Configure the Model 3155 as follows:
Function: Sine
Frequency:
1.000 kHz
Amplitude: 7.5 V
Remote Commands: :RES
:OUTP ON
:FUNC:SHAP SIN
:FREQ 1000
:VOLT 15
Summary of Contents for 3155
Page 4: ...This page was left intentionally blank...
Page 14: ...User Manual 3155 1 2 Overview Figure 1 1 3155 Series...
Page 41: ...User Manual 3155 Installation 2 15 Figure 2 15 Device Manager...
Page 44: ...User Manual 3155 2 18 Installation Figure 2 19 Customer Information Step...
Page 110: ...User Manual 3155 3 58 ArbConnection This page was left intentionally blank...
Page 150: ...User Manual 3155 Programming Reference 4 40...
Page 184: ...User Manual 3155 Programming Reference 4 74 This page was left intentionally blank...
Page 202: ...User Manual 3155 Specifications A 8 This page was left intentionally blank...