1-8 Overview
Front Panel
Connectors
The 3155 has 3 BNC connectors on its front panel: main and SYNC
outputs and trigger input. There are also 2 SMB connectors: sine
output and 10MHz reference input. These connectors are described
below.
Output
The output connector outputs fixed (pre-defined) waveforms up to
50MHz, user (arbitrary) and sequenced waveforms with sampling
clock to 100MS/s. Output impedance is 50
Ω
; the cable connected to
this output should be terminated with 50
Ω
load. Amplitude accuracy
is calibrated when connected to a 50
Ω
load. The amplitude is dou-
bled when the output impedance is above 1M
Ω
.
SYNC Output
The SYNC output generates a single TTL pulse for synchronizing
other instruments (i.e., an oscilloscope) to the output waveform. The
SYNC signal always appears at a fixed point relative to the wave-
form. The location of the SYNC signal along the waveform is pro-
grammable. The SYNC output is also used as a marker output when
the sweep or FM functions are turned on.
TRIG/FSK IN
In general, this input accepts signals that stimulate generation of
output waveforms. The trigger input is inactive when the generator
operates in continuous mode. When placed in trigger, gated or burst
modes, the trigger input is made active and waits for the right condi-
tion to trigger the instrument. In trigger and burst modes, the trigger
input is edge sensitive, i.e., it senses transitions from high to low or
from low to high to trigger the 3155. The direction of the transition is
programmable. In gated mode, the trigger input is level sensitive,
i.e., the generator is gated on when the logic level is high and idle
when the level is logic low. Trigger level for this input is TTL.
The same input is used in FSK mode, where the output hops be-
tween two frequencies – carrier and shifted frequencies. The output
generates carrier frequency when the FSK input is false and shifted
frequency when the FSK input is true.
The trigger input is also used as stop and start input when the 3155
is placed in Arm mode.
SINE OUT
This SMB connector outputs DC coupled, fixed level (1V
pk-pk
into
50
Ω
) sine waveforms. This output is derived directly from the sample
clock generator and is active at all times, regardless of present oper-
ating mode of the 3155. The frequency of the sine output is pro-
grammed using the sample clock parameter. Frequency agility and
Summary of Contents for 3155
Page 4: ...This page was left intentionally blank...
Page 14: ...User Manual 3155 1 2 Overview Figure 1 1 3155 Series...
Page 41: ...User Manual 3155 Installation 2 15 Figure 2 15 Device Manager...
Page 44: ...User Manual 3155 2 18 Installation Figure 2 19 Customer Information Step...
Page 110: ...User Manual 3155 3 58 ArbConnection This page was left intentionally blank...
Page 150: ...User Manual 3155 Programming Reference 4 40...
Page 184: ...User Manual 3155 Programming Reference 4 74 This page was left intentionally blank...
Page 202: ...User Manual 3155 Specifications A 8 This page was left intentionally blank...