13.2 Troubleshooting (480E PRO Only)
PROblEM
POSSIblE CaUSE
fIX
Unit will not power up
No voltage to Unit
Replace Power Supply
Unit does not beep
No voltage to Unit
Replace Switch
Pendant failure
Replace Pendant
Check pendant cable
at both ends
Return for service
Pendant, erratic display
Pendant failure
Replace Pendant
Damaged Display
Replace Pendant Display
Mechanical binding/jerking
Insufficient lubrication on track,
Use a Light Lithium
ballscrew, and track seals
based lubricant, Lubriplate #105
on tracks and ballscrews, and a
Silicone spray on track seals
Bearing bracket assembly failure Replace bearing bracket
Assembly
Ballscrew failure
Replace Ball Screw
Assembly
U-Bracket/Slider Assembly failure Replace U-bracket/Slider
Assembly Bracket
Assembly
Motor failure
Replace Motor
Return for service
Insufficient lifting power
Check that return on load (ROL) Increase force setting with
setting is sufficient
pendant controller
Note: For error codes see error codes in section 12.2.
Summary of Contents for 480
Page 1: ...480 480E 480E PRO CPM Service and Technical Manual 2011 QAL Medical 05070812 3 Rev C 08 11...
Page 2: ......
Page 25: ...25 14 Assembly Drawings 14 1 Main Assembly...
Page 26: ...26 14 2 Thigh Assembly...
Page 27: ...27 14 3 Cradle Sub Assembly 480 L480SA011 480E L480SA022...
Page 28: ...28 14 4 Foot Cradle Assembly...
Page 29: ...29 14 5 Ballscrew Assembly...
Page 30: ...30 14 6 480 U Bracket Slider Assembly...
Page 31: ...31 14 7 480E 480E PRO U Bracket Slider Assembly...
Page 32: ...32 14 8 480E or 480E PRO Motor Assembly Faulhaber...
Page 33: ...33 14 9 480E PRO ONLY Motor Assembly Maxon...
Page 34: ...34 14 10 Motion Controller Assembly 480 480E...
Page 36: ...36 14 12 480 Wiring Diagram...
Page 37: ...37 14 13 480E Wiring Diagram...
Page 39: ...39 14 15 480 480E Electrical Schematic...
Page 40: ...40...
Page 41: ...41...