12.2 Error Codes (480E PRO)
If an error occurs in the PRO Controller, a pop-up window will appear with the error code. A list of all error codes
appears below.
Error Code Table with Repair Steps
Error ID
Error Desc.
Repair Steps
01.01.01.00
Keypad
Keypad failure, replace keypad
01.01.03.xx
Memory Error
Reset machine, replace pendant
01.01.04.00
RTC Error
Reset machine, if persists, replace pendant
02.01.02.xx
Leading Pot
Reset machine, repair/replace pot, replace motor board
02.01.03.00
Raw pot value too large
02.01.04.00
Raw pot value too low
02.01.05.00
Hall Error
reset machine, replace motor, replace motor board
02.01.07.00
Actuator EEPROM Error
reset machine, replace motor board
02.01.08.xx
Communication Error
reset machine, replace cables, replace motor board, replace
pendant PCB
02.01.09.00
Too Many ROL's
reset machine, replace motor board, replace motor
02.01.10.01
Heartbeat Lost
reset machine, replace cables, replace motor board, replace
pendant PCB
Summary of Contents for 480
Page 1: ...480 480E 480E PRO CPM Service and Technical Manual 2011 QAL Medical 05070812 3 Rev C 08 11...
Page 2: ......
Page 25: ...25 14 Assembly Drawings 14 1 Main Assembly...
Page 26: ...26 14 2 Thigh Assembly...
Page 27: ...27 14 3 Cradle Sub Assembly 480 L480SA011 480E L480SA022...
Page 28: ...28 14 4 Foot Cradle Assembly...
Page 29: ...29 14 5 Ballscrew Assembly...
Page 30: ...30 14 6 480 U Bracket Slider Assembly...
Page 31: ...31 14 7 480E 480E PRO U Bracket Slider Assembly...
Page 32: ...32 14 8 480E or 480E PRO Motor Assembly Faulhaber...
Page 33: ...33 14 9 480E PRO ONLY Motor Assembly Maxon...
Page 34: ...34 14 10 Motion Controller Assembly 480 480E...
Page 36: ...36 14 12 480 Wiring Diagram...
Page 37: ...37 14 13 480E Wiring Diagram...
Page 39: ...39 14 15 480 480E Electrical Schematic...
Page 40: ...40...
Page 41: ...41...