Appendix
Guidance and manufacturer's declaration – electromagnetic emissions-
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration – electromagnetic emission
The device is intended for use in the electromagnetic environment specified below. The customer of the
user of the
device
should assure that it is used in such and environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The
device
uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emission
CISPR 11
Class B
The
device
is suitable for use in all establishments
other than domestic and those directly connected
to a low voltage power supply network which
supplies buildings used for domestic purposes.
Guidance and manufacturer's declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The
device
is intended for use in the electromagnetic environment specified below. The customer or the
user of
device
should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete
or ceramic tile. If floor are
covered with synthetic material,
the relative humidity should be
at least 30%.
Power frequency
(50/60Hz) magnetic
field
IEC61000-4-8
3A/m
3A/m
Mains power quality should be
that of a typical commercial or
hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration – electromagnetic immunity –
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the
user of device should assure that it is used in such an environment.
Immunity
test
IEC60601
test level
Compliance
level
Electromagnetic environment - guidance