EN
Select "SAVE" after completing time setup, time change is successful and exit system time setup and
return the previous menu. Select "CANCEL" to cancel the setting and return to the previous menu.
3.4.3 Data Delete
Select "YES" in "DELETE DATA" menu of [SYSTEM MENU], after you press certain key, the following
menu will pop up:
Figure 3.4.4 Data Delete
If you press "CONFIRM", the common user data will be deleted, press "CANCEL", the operation will be
canceled.
3.4.4 Prompt Setup
Select "PROMPT SET" item in [SYSTEM MENU] to enter its setup interface, then make corresponding
settings according to the following procedure:
"SYS PROMPT" and "DIA PROMPT" can control closing or unsealing of the SYS and DIA prompt
separately.
The prompt is on or off according to the high and low limits which have been set up. When the measure
result is higher than the high limit or lower than the low limit, and meanwhile, the "PROMPT SOUND" is
on, "SYS PROMPT" or "DIA PROMPT" accordingly on, the prompt will occur.
The adjustable ranges of the high and low limits of the adult mode prompt are as follows:
SYS PROMPT: 40~270 mmHg
DIA PROMPT: 10~215 mmHg
The adjustable ranges of the high and low limits of the pediatric mode prompt are as follows:
SYS PROMPT: 40~200 mmHg
DIA PROMPT: 10~150 mmHg
The adjustable ranges of the high and low limits of the neonatal mode prompt are as follows:
SYS PROMPT: 40~135 mmHg
DIA PROMPT: 10~100 mmHg
"DEFAULT" includes the main content:
Measurement mode: adult;
Parameter prompting limit:
User mode
High pressure
high limit
High
pressure
low limit
Low
pressure
high limit
Low
pressure
low limit
Adult
140
90
90
40
Pediatric
120
70
70
40
Neonatal
100
60
60
20
PROMPT SOUND switch: ON;
Measure unit: mmHg;
Ordinary user backlight time: 15s;
ABPM switch: END;
Asleep time: 22:00;
Asleep measurement interval: 30minutes;
Awake measurement interval: 15minutes;
Awake time: 7:00;
SYS PROMPT switch: ON;
DIA PROMPT switch: ON.
Note: The monitor does not have an alarm system, the above-mentioned "ALARM" represents
audible/visual prompt.
3.4.5 ABPM Menu
4.