EN
Figure 6.16.2 Measure Data Page Preview
Figure 6.16.3 Measure Data Page Preview
6.17 Help
Click shortcut key
to pop-up help file.It is the brief explanation for program function.
6.18 Exit
Click shortcut key
to save configuration information and exit application
7. Specification
Name
Ambulatory Blood Pressure Monitor
The degree of protection
against ingress of water
IP22
Display
2.4'' color LCD Display
Operating mode
Continuous operating
NIBP Specifications
Measurement Method
Oscillometric method
Working modes
Automatic
Measurement Range
Pressure
adult
0~290 mmHg(0~38.6 kPa)
SYS
:
40
~
270 mmHg
DIA
:
10
~
215 mmHg
pediatric
0~235 mmHg(0~31.3 kPa)
SYS
:
40
~
200 mmHg
DIA
:
10
~
150 mmHg
neonatal
0~140 mmHg(0~18.6 kPa)
SYS
:
40
~
135 mmHg
DIA
:
10
~
100 mmHg
Pulse: 40~240/min
Inflation
Adult
160±5 mmHg (21.33±0.67 kPa)
pediatric
120±5 mmHg (16±0.67 kPa)
neonatal
70±5 mmHg (9.33±0.67 kPa)
Prompt Range
adult mode
SYS PROMPT: 40~270 mmHg
DIA PROMPT: 10~215 mmHg
pediatric mode
SYS PROMPT: 40~200 mmHg
DIA PROMPT: 10~150 mmHg
neonatal mode
SYS PROMPT: 40~135 mmHg
DIA PROMPT: 10~100 mmHg