3.9
Trace
The
AE-366B
is able to set the parameters of up to 3 different traces on the
display at once. Each trace is represented by a different color and is updated with
each sweep.
3.9.1
Selecting a Trace
Each trace (A, B, C) is represented by a different color. Trace A is green, trace B
is orange and trace C is yellow. When activated, an icon for each trace color and
function is shown at the bottom of the display. When a trace is selected,
parameters can be set/edited from the Trace menu.
Figure 47.
►
Trace Type & Icon
The type of trace used determines how the trace data is stored or manipulated
before being displayed. The analyzer updates each trace according to the type
of trace used.
Clear & Write
:
The
AE-366B
continuously updates the display with each sweep. This is
the default trace type.
Peak Hold,
Min Hold
:
The maximum or minimum points are maintained for the selected trace.
The trace points are updated each sweep if new maximum or minimum
points are found.
View
:
View will hold the selected trace and stop updating the trace data for the
selected trace. Pressing
View
will display the trace data that was
cleared using the
Blank
key.
3-46
July 2018