July 2018
3-35
3.6
Measurement
This section describes how to use the automatic measurement modes. The
AE-366B
includes the following measurements:
ACPR.
OCBW.
3.6.1
Channel Analysis Overview
Channel analysis measurement includes ACPR (adjacent channel power) and
OCBW (occupied bandwidth) measurements.
►
Parameters
:
Channel bandwidth
:
The frequency bandwidth the target channel occupies.
Range: Between 0 Hz ~ 3 GHz (0 Hz excepted).
Channel Space
:
The frequency distance between each main channel.
Range: Between 0 Hz ~ 3 GHz.
Adjacent channel bandwidth 1 & 2
:
The frequency bandwidth the adjacent channels occupy.
Range: Between 0 Hz ~ 3 GHz (0 Hz excepted).
Adjacent channel offset 1 ~ 2
The frequency distance between the adjacent channels and main channel.
Range: 1 Between 0 Hz ~ 3 GHz (0 Hz excepted).
OCBW%
The ratio of occupied bandwidth to the amount of power consumed.
Range: 0 % to 100 %, 0,1 % resolution.