July 2018
3-43
Use the arrow keys to move the cursor to the frequency column of the
desired point.
Figure 45.
Enter the new frequency and amplitude of the point using the keypad and
the unit keys.
Repeat steps 3-5 for the remaining points (A maximum of ten points).
To delete the selected point, press Delete
.
To delete all the points, press Delete All…
.
The points will revert to their default frequency and amplitude
values.
To delete a point from the editing table, press Delete.
The whole points, including the frequency value and the
amplitude are removed. This command reduces the number of
points used in the limit line.
Press
Undelete
the restore the last point that was deleted.
3.7.2.1
Pass/Fail Testing
The Pass/Fail testing uses the limit lines as boundaries. When the input signal
escapes the boundary of the limit lines, then the test is judged as a FAIL, if the
signal stays within the boundary, the test is judged as a PASS.
Figure 46.