4-6 PRODIGIT
4. Please make sure range I/II command before execute the load current setting command,
otherwise the 33501 Series load current after programming the RANGE command.
5. LOW|HIGH option is for 33501 series electronic load module.
6. The engineering unit for load resistance is Ohms.
For Example
:
RES
:
LOW 0.123 set LOW level load resistance to 0.123 OHM.
RES
:
HIGH 3.456789 set HIGH level load resistance to 3.456789 OHM.
LOAD
Purpose
:
Turn the Electronic load module input ON or OFF.
Command syntax
:
LOAD {SP}{0FF|ON}{NL}
Description
:
This command sets the electronic load to sink current from DC power source.
LOAD OFF; the 33501 Series Electronic load at input OFF condition.
LOAD 1; set to LOAD ON status, this load is ready to sink current from power source.
LEV
Purpose
:
Select Low or High level in static mode.
Command syntax
:
[GLOB
:
] LEV {SP} {0|1|LOW|HIGH} {NL}
Description
:
LEV LOW ; Set LOW current level in CC mode, LOW resistance level in CR mode, or LOW
voltage level in CV mode, LOW resistance level in CP mode.
LEV 1 ; Set HIGH current level in CC mode, HIGH resistance level in CR mode, or HIGH
voltage level in CV mode, HIGH resistance level in CP mode.
PRES
Purpose
:
Set the upper or lower 4 1/2 digit multi-function meter to display the programming
load level.
Command syntax
:
PRES {SP} {0|1|OFF|ON} {NL}
Description
:
PRES ON set the load to preset on status.
MODE
Purpose
:
Select the operating mode of electronic load module.
Command syntax
:
MODE {SP} {0|1|2|3|CC|CR|CV|CP} {NL}
Description
:
MODE CC; set the presently operating mode to Constant current mode.
MODE CV; set the presently operating mode to Constant Voltage mode.
MODE CR; set the presently operating mode to Constant Resistance mode.
MODE CP; set the presently operating mode to Constant Power mode.