33501 Series Operation Manual 2-7
2-14 AUTO SEQUENCE testing function description
There are two modes in AUTO SEQUENCE function, EDIT MODE and TEST MODE, The
AUTO SEQ mode can be entered by press S3 + S4 key simultaneously, then press STORE
key again to enter the EDIT MODE, or press START key to enter the TEST MODE, Please
refer to the flow chart operation below:
Normal Mode
AUTO SEQ.
Press S3+S4 key
Display bank no.
Press Exit key
bn
Show r1-r9
Display 3300C Firmware version r1-r9
Display bank
STATE MODE
STORE MODE
Press STATE key
Press STORE key
Show n1~n9
Fig 2-8 AUTO SEQ function operation flow chart
2-14-1 EDIT MODE
The TEST Mode of Auto-Sequence function is entered by press S3 + S4 key
simultaneously, The EDIT mode of Auto Sequence function is entered by press S3 +
S4 key simultaneously, the S3 and S4 LED are ON to indicate Auto-sequence mode,
then the Edit mode of Auto-sequence function is proceed by press STORE key.
The EDIT MODE flow chart is described below:
2-14-1-1 There are nine Auto Sequence (n1-n9) can be edit within 33501 series.
2-14-1-2 Each Auto Sequence has up to 16 Test step, where each step is any one
memory of 150 sets Store memory which is 30 memory Bank with 5
memory state.
2-14-1-3 Each test step has t1 (test time) and t2 (delay time), the unit is 100mS,
the range is 0.1S - 9.9S in 100mS resolution. 33501 series mainframe will
check each module GO/NG at the end of t1 (test time), the next step will
be started after duration t2 (delay time).
2-14-1-4 The test step sequence can be up to 16 step, and can be terminated by
press EXIT key if less than 16 step is required.