33501 Series Operation Manual 3-17
Table 3-11 33533 initialize
3-3 Input terminal and wire consideration
The five ways connect the input wires to the Electronic load the connection methods are
made as follow:
3.3.1
Plug connectors: This is the most popular way to connect the input of
electronic load to the device under test. It is recommended the load
current is less than 20A in this connection for the current rating of the plug
is rated to 20A. The maximum wire gage AWG14 can be used in this
application.
3.3.2
Spade terminals: The spade terminal provides a good contact to the
binding post, it is recommended to use anytime. The maximum wire gage
10 can be used in this application.
3.3.3
Insert the wire into the input terminal: This is the most convenient way to
connect the load input and D.U.T. The maximum wire gage AWG14 can
be used in this application.
3.3.4
Both plug connectors and spade terminals:
It is recommended to use when input current is greater than 20A or long
lead wires.
3.3.5
Both plug connectors and Insert the wire into the input terminal.
It is recommended to use when input current is greater than 20A or long
lead wires.
A major consideration in making input connection is the wire size. The
minimum wire size is required to prevent overheating and to maintain
good regulation. It is recommended that the wires should be large enough
項
目
起始值
項
目
起始值
CC L
0.000 A
VH
60.00 V
CC H
0.000 A
VL
0.000 V
CR H
234.3
Ω
AH 960.0
A
CR L
234.3
Ω
AL 0.000
A
CV H
60.00 V
WH
7200.0 W
CV L
60.00 V
WL
0.000 W
CP L
0.000W
SENSE
Auto
CP H
0.000W
LD-ON
0.1 V
FALL
4000 mA/uS
LD-OFF
0.0 V
RISE 4000
mA/uS
RANGE AUTO
T HI
0.05 mS
T L0
0.05 mS