33501 Series Operation Manual 3-3
3.1.4
Remote LCD Indicator
The Remote LCD Indicator is used to indicate the status of remote operation, all of
the front panel operation can not be operated while Remote LCD is ON, in case of
Local mode or manual operation, the Remote LCD is OFF.
3.1.5
Upper 5 digit LCD display
The 5 digit LCD display is a multi-function display, the functions are described below:
Normal mode :
There is a 5 digit DVM display, display measuring data of the DC input terminal or V-
sense input terminal if V-sense AUTO is programmed, or the 5 digit voltage meter
displays the voltage of V-sense input terminal if V-sense ON is programmed.
When the auto-sense of V-sense function is programmed, the auto-sense circuit of
33501 series electronic load can check the V-sense cable is connected or not, the V-
sense BNC input is detected if it is greater than 0.5V (33501, 33511) or 2.5V (33521)
or not, if yes then the 5 digit DVM measures the sense BNC input, otherwise, the 5
digit DVM measures the DC input terminals of the load module.
Test Setting Mode:
Short : Short test Enable and Short Setting programming
:
Display will show “Short”.
OPP : OPP test Enable and OPP Setting programming
:
Display will show”OPP”.
OCP : OCP test Enable and OCP Setting programming
:
Display will show”OCP”.
Short testing
、
OCP testing and OPP testing programming, will show Vsense’s
voltage or load Input voltage.
3.1.6
Middle 5 digit LCD display
Normal mode:
There is a 5 digit DAM display. The 5 digit DAM displays the measuring current of the
DC load When Load ON programming.
Setting Mode:
3.1.6.1 Config ON programming: Display will individually show”SENSE”, “LDon”,
“LDoff” and “POLAR”.
3.1.6.2 Limit ON programming
:
Display will individually show”V_Hi”, “V_Lo”,
“A_Hi”, “A_Lo”, “W_Hi”, “W_Lo” and “NG”.
3.1.6.3 DYN setting ON programming
:
Display will individually show ”T-Hi”, ”T-Lo”,
”RISE” and “FALL”.
3.1.6.4 Short test Enable
、
OCP test Enable and OPP test Enable programming:
Display will individually show
「
PRESS
」
.
3.1.6.5 Short setting programming
:
Display will individually show “TImE”, “V-Hi”
and “V-Lo”.
3.1.6.6 OPP setting programming
:
Display will individually show “PSTAR”,
“PSTEP”, “PSTOP” and “Vth”.
3.1.6.7 OCP setting programming
:
Display will individually show “ISTAR”, “ISTEP”,
“ISTOP” and ”Vth”.
3.1.6.8 Short testing programming, the current of actual load current, the unit is "A".
3.1.6.9
OCP testing programming, the seting current , the unit is "A".
3.1.6.10 OPP testing programming, the seting watt , the unit is "A"
3.1.6.11 When Over current protect
:
Display will show
「
OCP
」
.