82
PI-MAX
®
3 System Manual
Version 1.A
Figure 62. Trigger In and Trigger Out tabs
6.
When the experiment is ready, click on the
ACQ
button or select
Acquire
on the
Acquisition
menu to start the image acquisition.
Tips and Tricks
Experiments using the DIF feature of the PI-MAX3 can be complex, and timing of the
events is usually rather exacting. Here are several points to consider that may make the
experiment setup or troubleshooting much smoother and easier.
The most important piece of equipment in a DIF experiment is an oscilloscope. The
PI-MAX3 has a
MONITOR
BNC on the back of the camera which is very useful for
seeing when the two image exposures occur during the course of the experiment. The
use of the
MONITOR
BNC and an oscilloscope is discussed in more detail in
Chapter 9 - Tips and Tricks
.
The short time between the two images in DIF requires an intensifier with a fast
phosphor. P46 phosphor has a decay time of ~ 2 µs which means it takes 2 µs for the
phosphor emission to drop to 10% of its peak value. The decay is not a simple single
exponential; even after 100 µs there may be 1% or more of the first image on the
phosphor screen. It is usually possible to subtract a percentage of the first image from
the second image to remove the residual image. If this is not possible, there are
intensifiers with P47 phosphor, which is an order of magnitude faster than P46.
The software uses the Phosphor Decay Time to determine how long to wait after the
gate pulse to shift the image. This value can be adjusted in the Hardware Setup
dialog. If there is some residual image from the first frame in the second frame,
simply increase the Phosphor Decay Time to allow the phosphor more time to decay
before shifting the image. If residual image is not an issue, then Phosphor Decay
Time can be shortened to decrease the time between the two DIF images.
Summary of Contents for PI-MAX 3 System
Page 1: ...4411 0129 Version 1 A September 8 2010 4411 0069 ...
Page 32: ...32 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 52: ...52 PI MAX 3 System Manual Version 1 A Figure 20 Safe Mode and Fast Mode Operation ...
Page 72: ...72 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 90: ...90 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 100: ...100 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 110: ...110 PI MAX 3 System Manual Version 1 A Figure 68 Outline Drawing PI MAX3 with F mount Adapter ...
Page 114: ...114 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 120: ...120 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 140: ...140 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...