76
PI-MAX
®
3 System Manual
Version 1.A
Timing Modes
In WinView and WinSpec, the timing modes available in the
Acquisition|Experiment
Setup…|Timing
tab are different from those in standard intensified systems. The
available timing modes are:
Single Trig. Mode:
two shot, one trigger for both shots.
Dual Trig. Mode:
two shot, each shot requires a trigger.
The trigger(s) can be generated by an external source connected to the PI-MAX3 or the
PI-MAX3 can generate the trigger(s) internally.
Setup
Hardware
Figure 48. System Diagram: DIF Operation
Software
For the purposes of this setup, it is assumed that you are using either WinView or
WinSpec to control the system.
Operation
The operation of the PI-MAX3 in DIF mode is similar to the standard operation of a
PI-MAX3 with SuperSYNCHRO. There are only a few differences due to the special
timing modes of DIF, and they will be outlined here. Because there are two timing modes
in DIF operation, there are two procedures for setting up the experiment.
Summary of Contents for PI-MAX 3 System
Page 1: ...4411 0129 Version 1 A September 8 2010 4411 0069 ...
Page 32: ...32 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 52: ...52 PI MAX 3 System Manual Version 1 A Figure 20 Safe Mode and Fast Mode Operation ...
Page 72: ...72 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 90: ...90 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 100: ...100 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 110: ...110 PI MAX 3 System Manual Version 1 A Figure 68 Outline Drawing PI MAX3 with F mount Adapter ...
Page 114: ...114 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 120: ...120 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 140: ...140 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...