Dual Trigger Mode
Figure 56. DIF Operation: Dual Trigger Timing Diagram
1.
The first requirement is that the PI-MAX3
camera be aligned and focused on the area
of interest in the experiment. This is best
accomplished while the PI-MAX3 is
operating in Interline mode (i.e., before
switching to DIF mode). Verify that the
Phosphor Decay Time
is appropriate to
the phosphor used by your camera: (this
field is located on
the
Setup| Hardware
Setup|Controller/ Camera
tab). The
procedure for initial focus is outlined in
Chapter 4.
2.
After the alignment and focus, the
PI-MAX3 system needs to be put into DIF
mode. On the
Setup|Hardware
Setup|Controller/Camera
tab, select
Dual Image Feature
as the
Readout
Mode
, and then click
OK
.
Figure 57. Hardware
Setup|Controller/Camera tab
Summary of Contents for PI-MAX 3 System
Page 1: ...4411 0129 Version 1 A September 8 2010 4411 0069 ...
Page 32: ...32 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 52: ...52 PI MAX 3 System Manual Version 1 A Figure 20 Safe Mode and Fast Mode Operation ...
Page 72: ...72 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 90: ...90 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 100: ...100 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 110: ...110 PI MAX 3 System Manual Version 1 A Figure 68 Outline Drawing PI MAX3 with F mount Adapter ...
Page 114: ...114 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 120: ...120 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 140: ...140 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...