Attention!
Intensified CCD detectors, such as the PI-MAX
®
3,
when biased ON, can be irreparably
damaged if continuously exposed to light levels higher than twice the A/D saturation
level. Thus it is
critical
that you
not
establish conditions that could result in damage to
the intensifier. Although intensified detectors are less prone to damage from
background
light
when operated gated, they are at
significant risk
to damage from high-intensity light
sources like a laser. High intensity sources can damage the intensifier before the
protection circuits have time to respond, or even cause spot damage without the
protection circuits acting at all. If a sustained alarm indication occurs when the controller
is turned on, either completely cover the intensifier to reduce the light to halt the overload
condition, or reduce the laboratory illumination still further until safe operating
conditions are established.
Alarm
To reduce the risk of detector damage, the PI-MAX3
detector is equipped with an audible
alarm in the detector head, activated when the intensity of light falling on the image
intensifier exceeds a preset threshold. While the alarm is sounding, the photocathode is
disabled. Immediately switch the
I.I.T.
switch (on the back of the PI-MAX3) to the
OFF
position. Cover the detector window and only switch the
I.I.T.
switch to
ON
after the
illumination level has been lowered. If the alarm sounds continuously even when the
illumination level is adequately low, shut the system down and contact the factory for
guidance.
Note:
It is normal for the alarm to sound briefly when the system is turned on.
Discontinue operation and contact the factory at once if sporadic or continuous
unwarranted alarms occur. They may indicate intensifier damage or another situation that
requires immediate attention.
Unigen
TM
II Coating
For greater sensitivity, the PI-MAX3 camera with Unigen II coating comes without a
protective window over the intensifier. Do not scratch the image intensifier’s coated
surface. To remove any lint or dust use very clean dry nitrogen with < 5 PSI pressure. If
you have any questions please contact your sales contact or the factory.
WARNING
Caution
Summary of Contents for PI-MAX 3 System
Page 1: ...4411 0129 Version 1 A September 8 2010 4411 0069 ...
Page 32: ...32 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 52: ...52 PI MAX 3 System Manual Version 1 A Figure 20 Safe Mode and Fast Mode Operation ...
Page 72: ...72 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 90: ...90 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 100: ...100 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 110: ...110 PI MAX 3 System Manual Version 1 A Figure 68 Outline Drawing PI MAX3 with F mount Adapter ...
Page 114: ...114 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 120: ...120 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 140: ...140 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...