Single Sequence
A sequence is the repetition of one pulse set “X” number of times. Figure 46 shows a
sequence with the pulse set repeated three (3) times.
Figure 46. Sequence with 3 Repetitions
The EXPOSE signal remains high for the duration of the sequence and goes low
approximately 15 ns after sequence completion.
Supported Timing Generator Trigger Modes:
Trigger per Pulse:
Each pulse of the single sequence is initiated by its own trigger (all
triggers must be from the same source and all must be either internal or external).
See
Supported Trigger Sources:
Source 1: Timing Generator internal trigger
. Each pulse set of the sequence
is initiated by the timing generator’s internal trigger.
Source 2: External Trigger (on TRIGGER IN BNC)
. Each pulse set is
initiated by an external trigger.
Figure 47. Timing Diagram: Sequence with 3 Repetitions and Trigger per Pulse
Summary of Contents for PI-MAX 3 System
Page 1: ...4411 0129 Version 1 A September 8 2010 4411 0069 ...
Page 32: ...32 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 52: ...52 PI MAX 3 System Manual Version 1 A Figure 20 Safe Mode and Fast Mode Operation ...
Page 72: ...72 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 90: ...90 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 100: ...100 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 110: ...110 PI MAX 3 System Manual Version 1 A Figure 68 Outline Drawing PI MAX3 with F mount Adapter ...
Page 114: ...114 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 120: ...120 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 140: ...140 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...